Highlights

  • Maximum Test Port Power: -2 dBm (Typ.)
  • Modulation Signal Path
    • Loss: 6 dB (Typ.)
    • Input Power P0.1dB / Damage: +10 dBm / >+20 dBm (Typ.)
  • Dynamic Range @IFBW = 10 Hz: 110 dB (120 dB Typ.)
  • Raw Directivity: 30 dB (Typ.)
  • Stability: ±0.25 dB, < ±4 deg (Typ.)
  • 1.2m or 5m cable set (optional)

The N5262CW05 is a frequency extender module supplied by VDI to perform measurements with a vector network analyzer in 140 to 220 GHz (G-band, WR05). It can be used with a PNA or PNA-X with an N5292A test set controller or directly connected with a 4-port PNA/PNA-X. It can also be connected with a 4-port Streamline VNA or PXI-VNA with an N5252APXI Test Set Adapter.

1 Extended Frequency Coverage applies; module performance within the standard band conforms to standard specifications while performance in the extended regions can be degraded as follows:

The minimum dynamic range is 95dB (15dB degradation compared to the standard band specification)

The test port power typical across the extended band is degraded by 5 dB or less compared to the specification for the standard band.

For more information, please refer to the technical overview, Banded Millimeter-wave Network Analysis, and visit VDI for specification tables and more information.