Situation:

The Baseband Generator board with part number E4400-60070, serial number starting with 300, and version number A0110 contains wrong header information. This condition generates I/Q offset problems. To check if your E443xB ESG-D has the defect, access the installed board information on the display of the ESG. Using the front panel press Utility > Instrument Information > Installed Board Information.

Solution:

To remedy the problem, download and run the self extracting executable Software Patch "BBG_rev.exe" available below.

是德建议使用最新的版本,其中包括最新的漏洞修复和安全补丁
发布日期 版本 版本说明
2004-02-20 1.0 Self-extracting executable Software Patch.
发布日期 版本
2004-02-20 1.0
Version Description
Self-extracting executable Software Patch.

如何下载

  1. Click the Download button
  2. Save the file to your PC

如何安装

  1. Activate the executable file
  2. Follow the installation prompts

颁发许可证

  • Does Not Require a License

Installs on: Instrument

前提要求

  • None

支持的仪器

  • E4430B, E4431B, E4432B, E4433B, E4434B, E4435B, E4436B, E4437B with Option UN8 or UN8 UN9.
没有可用的软件版本
SHOW MORE